Phantom Pulse Calculator
CTH:YAG Phantom Pulse Calculator
Estimate the phantom pulse parameters required to maintain the thermal
loading associated with the optimum CTH:YAG laser operating condition.
Laser Parameters
Electrical pump energy of the normal lasing pulse.
Repetition rate at which the resonator reaches the desired
thermal operating condition.
Phantom Pulse Parameters
Used with the MWL CTH:YAG standard-test calibration to estimate
pump energy.
Enter ONE Phantom Pulse Parameter
Enter either pulse width or repetition rate and leave the other
field blank. The missing value will be calculated.
OR
MWL CTH:YAG Calibration
Phantom pulse energy is estimated from MegaWatt Lasers
CTH:YAG 4 × 125 mm standard-test voltage and pump-energy data.
Intermediate voltages are determined by interpolation.
Calibration & Extrapolation Caveat
The voltage-to-pump-energy relationship used by this calculator is
based on MegaWatt Lasers CTH:YAG 4 × 125 mm standard-test data.
Pulse-width calculations outside the directly measured standard-test
range are extrapolations and should be treated as engineering estimates.
In particular, short phantom pulses substantially below the measured
test pulse widths may not follow the same energy-versus-pulse-width
relationship exactly because actual flashlamp discharge behavior
depends on the laser driver, flashlamp impedance, capacitor bank,
inductance, pulse-forming network, cooling conditions, and other system
parameters.
Calculated phantom-pulse settings should therefore be used as a
starting point and experimentally verified before use.
Calculated Phantom Pulse Requirements
Target Average Pump Power
—
Laser Average Pump Power
—
Required Phantom Average Power
—
Estimated Phantom Pulse Energy
—
Phantom Pulse Voltage
—
Phantom Pulse Width
—
Phantom Repetition Rate
—
Phantom Duty Cycle
—
Laser and Phantom Pulse Timing
One laser repetition period is shown. Pulse spacing is based on
repetition rate. Pulse widths are displayed using a compressed visual
scale so that microsecond pulses remain visible.
Laser Period
—
Phantom Period
—
Phantom Pulses Between Laser Pulses
—
Phantom Duty Cycle
—
Pulse spacing is shown using the actual repetition-rate relationship.
Pulse widths use a compressed visual scale because microsecond pulses
would otherwise be too narrow to see relative to a millisecond-scale
repetition period.
Engineering note:
This calculator estimates the phantom pulse conditions required to reproduce
the average pump loading associated with the resonator’s selected optimum
operating point. Actual optimum settings may differ because thermal lensing,
flashlamp coupling, rod cooling, resonator alignment, and electrical
discharge characteristics vary between systems.